Temperature Controller Accessories
Download
flyer
BT-1-xx Tissue chambers
Note: all current tissue chambers have tapered inlet area.

BT-1

BT-1-AR

BT-1-13

BT-1-13B

BT-1-RR

BT-1-18
BT-1-TB

BT-1-TBS
BT-1-TBS with STIM-TB
-
- Application
Tissue chambers for study of isolated cells, small pieces of tissue or brain
slices.
- size platform 8.3 x 5.1 x 0.55cm
- size central chamber
- BT-1: 22.4 x 8 x 1.3mm [LxWxD], approx. 0.25ml volume, bevelled edges
- BT-1-AR 16 x 5.5 x 1.3mm, approx 0.125ml, small volume chamber
- BT-1-12A: 22.4 x 12 x 3.5mm, approx. 0.45ml volume, partly bevelled
edges
- BT-1-13: approx. 0.5ml volume, BT-1-13B has bevelled edges
for lower solution levels and tapered inlet.
- BT-1-18: 22.4 x 18 x 3.5mm, approx. 1.4ml volume
- BT-1-RR: 28x11x1.3mm
- BT-1-TB: 16.5x14x1.3mm approx 0.5ml for 12mm cover slides
- BT-1-TBS: 16.5x14x1.3mm approx 0.5ml for 12mm cover slides, with field stimulation with STIM-TB
- slit type inflow & outflow for laminar flow, thermistor groove
- constructed of polycarbonate
- compatible with HI-25, HI-25T
- Don't see a chamber that is suitable. Call us about modifying our existing
chambers to your requirements. Depending on the changes it may not cost any
more than a standard chamber.
SA-LTEK2 LabTek Stage heater for LabTek chambers (2 window)
- Application
Heating LabTEK chambers. Use with inverted or conventional
microscope.
- resistance nominally 3.3ohms
- plugs into Heater Output of TC2bip
HI-25 Heater-ITO transparent

Download flyer
- Application
Heating BT-1-xx Tissue chambers or similar small
tissue chambers. Use with inverted or conventional
microscope.
- transparent indium tin oxide coated heater
- size 22x50x approx 0.25mm thick.
- Heated area approx:18x45mm
- resistance nominally 10ohm
- max. operating temperature 70C
- plugs into Heater Main of TC2bip,
TC-1
HI-25T/HI-24T Heater-transparent thin
Download
flyer
- Application
as for HI-25 but thinner for fluorescence (see transmission data)
and/or microscopy studies.
- HI-25T same as HI-25 except 0.14mm thick
- HI-24T same as HI-25T except 40mm wide,
heated area 18mmx35mm and with a resistance of 12.5ohms.
HI-55D/HI-57D Heater-transparent thick
Download flyer
- Application
heating 30mm culture dishes, Nunc Lab-Tek chambers.
Heater forms a platform
- for dishes/chambers.
- transparent indium tin oxide coated heater
- size: 50x50(70)x»0.6mm thick.
- resistance nominally 9ohm
- max. operating temp. 70°C
- plugs into Heater Main of TC2bip , TC-1
HLS-1P / HLS-8x.8P / HLS-8x1.3P Heater assy-lens, silicon
- Application
objective heater and thermistor probe. heater size 20x70x approx 0.35mm thick.
thermistor as for TH-10Km plugs into TC2-Ch3 also available
as heater alone with banana plugs HLS-1. Heater dimensions: HLS-1P: 7x2cm
HLS-8x.8P: 8x0.8cm HLS-8x1.3P: 8x1.3cm
TC2-Ch3 3rd Heat only channel

- Application
optional channel for TC2bip. Independent
controller.
- power from TC2bip max. 8V, 1.5Amp
output
- fits inside TC2bip
- user installable
HC-1 Heater-concentric foil
- Application
Heating 30mm culture dishes. Use with inverted or conventional
microscope.
- size 40mm outer dia. x 24mm inner dia.
- resistance nominally 6.4ohm
- constructed of a thin foil coated with Kapton insulation.
Total thickness
- 0.35mm. One side coated with pressure sensitive adhesive.
- max. operating temperature 100oC
- plugs into Heater Main of TC2bip,
TC-1
RW-10 (i) Resistance wire
- Application
Capillary pre-heater or other special purpose heaters
- RW-10: 3.3ohm/10cm uninsulated NiCr wire
- RW-10I: 2.6ohm/10cm insulated NiCr wire
HPRE/HPRE2 Pre-heater


Download flyer
- Application
3 loop capillary heater for pre-heating solution
- HPRE/HPRE2 dims. 45x14x7mm (LxWxH) heater resistance approx 11ohm,
50ul dead volume. Suitable for flow rates from 0 to at least 4.5ml/min.
- Thermistor sensor HPRE: 100ohm NTC HPRE2: 10kohm NTC
- HPRE plugs into HPREC Control box
- plugs into Heater Aux of TC2bip
MPRE8 Multitube Pre-heater

Download flyer
- Application
superfusion of cells or small tissue pieces
- Dims. 10x0.3cm (LxDiam); heater resistance approx 11ohms.
- Tip dead volume 1-2microl. Suitable for flow rates up to
1ml/min.
- Thermistor sensor: 10kohm NTC
- Plugs into Heater Aux or TC2-Ch3 of TC2bip
- Can use with SYRHT8K Syringe barrel heater kit to
degass/pre-heat solutions
HPREC Control box (TC-1 accessory)
- Application
Control box for HPRE capillary pre-heater
- HPREC used with TC-1 with flow rates
>2ml/min.
- plugs into B output of TC-1.
TH-2Km Interchangeable thermistor probe
- Application
mini 400 series compatible probe (approx. 0.45mm diam.).
- suitable for shallow baths etc.
- rugged alternative to the YSI-511
- Rth @ 25oC: 2225ohm
- time constant max operating temp: 100oC; 1m
lead; 6mm plug
- compatible with Main Probe input of TC2bip
TH-10Km Interchangeable thermistor probe
- Application
miniature thermistor probe (approx. 0.45mm diam.).
- general purpose mini probe, submersible
- Rth @ 25C: 10kohm
- time constant max operating temp: 100C; 1m lead; 6mm plug
- compatible with Aux Probe input of TC2bip
TH-1 Thermistor probe
- Application
miniature thermistor probe (approx. 0.35mm diam.).
- suitable for shallow baths , point temperature
measurement, submersible
- rugged alternative to the YSI-511
- compatible with YSI 43T series telethermometer Rth @ 25oC:
2kohm+/-5%; beta approx 3000
- time constant 20ms, plunge water
- max operating temp: 100oC; 2m lead; 6mm plug
- compatible with Main Probe of TC2bip
FL-1 Miniature positioner
- Application
Adjusting bath solution levels, positioning ref.
electrodes
- size: base 13mm dia., overall height 24mm
- adjuster height range approx 3mm; approx 0.5mm/turn
- miniature block for clamping tube, electrode
- spare spring wire for custom bending
TCAL Thermistor calibration software
- Application
Accurate thermistor calibration. Using several test
measurements (temperature, resistance) program computes
the relationship between the temperature indicated on a
YSI-Telethermometer and the actual temperature as well as
Rth versus temperature.
Last updated 19Dec00