Cell MicroControls, P.O. 11387 Norfolk, VA 23517 Ph: (800) 398-4CMC Fax: (800) 398-9CMC Email: info@cellmcl.com


TC2bip-Controls

Front controls
TEMP1,2:
setpoints allow switching between two preset temperatures.
POWER:
Controls overall output power to limit rate of temp increase and put controller in standby mode.
Aux/Main:
Display temperature of main probe or auxiliary probe. Output at rear can be used to monitor Aux temperature continuously with a DVM.
Cool/Heat or Heat:
Cool/Heat is used for Peltier devices. Heat is used for resistance type heaters.
OUTPUTS:
LEDs display output current and direction of current.
Aux On/Off:
Shuts down Aux output.
Low Pwr:
Indicates when power input is too low for proper function.

Rear controls
PROBE Main:
Accepts TH-2Km, YSI-400 and YSI-500 series or compatible probes. Two adjustments allow customization for miniature non-interchangeable probes (eg. TH-1).        
OUT
: Temperature output for DVM, computer or chart recorder.        
Ext In: Allows input from other probes (eg. Sensortek). Automatically switches to external input.
PROBE Aux:
Accepts inexpensive interchangeable thermistor probe.
T Aux Offset:
Offsets Aux reference +/-6oC.
LIMITS:
Sets max output voltage for particular heater type. Min is max output voltage for Peltier for cooling.
OUTPUTS:
Output to Main heater (or Peltier) and Aux heater (pre-heater or Peltier). (3rd channel uses MiniDIN connector for heater/thermistor.)
Fuses:
Protects controller power amplifier from overload.
Ext T Ref:
External temperature programming (eg. for computer setting of temperature).
PMain:
Proportional gain for main output.
IMain:
Integral gain for main output.
Integ On
: Integrator on/off control.
PAux:
Proportional gain for auxiliary output.
GMain:
Overall gain of output amplifier for main output.
GAux:
Overall gain of output amplifier for auxiliary output.
Aux=Main
: Aux output=GAux/GMain x Main output (pre-heater as slave).


Last updated 9 Jan 97
© copyright 1995-2000 N. B. DATYNER